Pacing
Cardiac pacing is repetitive, low-current electrical stimulation of the heart to initiate and maintain rhythm. Pacing:
- Indications include:
- Bradycardias
Generally symptomatic heart block or sinus node dysfunction, or asymptomatic with HR <40. - Anti-tachycardia therapy
Over-drive pacing of SVT (AVNRT, AVRT) and VT. - Resynchronisation therapy
↑ EF by coordinating LV and RV contraction.
- Bradycardias
- May be performed:
- Transvenous
- Permanent
Classic insertion of pacing leads via the SCV into the RA/RV - Temporary
Flotation of a pacing catheter via a sheath for temporary emergency pacing.
- Permanent
- Epicardial
- Temporary
Temporary epicardial wires inserted during cardiac surgery. Usually deteriorate over 5-10 days. - Permanent
Occasionally permanent epicardial wires may be inserted during cardiac surgery (or via VATS) if venous access is compromised.
- Temporary
- Transcutaneous
Emergency pacing via pads. Requires sedation.
- Transvenous
Terminology
Key pacing terminology includes:
- Sensitivity
- Thresholds
- Undersensing
- Oversensing
- Capture
- Thresholds
- Pacemaker code nomenclature
- Pacing modes
Sensitivity
Sensitivity describes the ability of the pacemaker to detect native cardiac rhythm. Sensitivity is:
- Measured in mV
- The sensing threshold is the voltage at which native activity is detected
- The sensitivity setting (usually known as sensitivity) is the voltage that the pacemaker is set to
- This is lower than the sensing threshold
“Increasing the sensitivity” therefore requires decreasing the sensitivity setting.
- This is lower than the sensing threshold
- Good when the sensing threshold is high
- Poor when the sensing threshold is low
Undersensing:
- Failure to sense the normal cardiac rhythm, resulting in inappropriate delivery of a pacing impulse
If this occurs during chamber repolarisation, arrhythmia can result:- Inappropriate atrial pacing may cause AF
- Inappropriate ventricular pacing may cause polymorphic VT or VF
- May occur due to:
- Variability in the sensing threshold
- Inappropriate setting of the sensitivity
- Due to the risk of malignant arrhythmias, use undersensing pacing leads should be carefully considered and not used at all outside of a critical care setting
Low atrial sensitivity is less concerning than low ventricular sensitivity as there is less (but not no) R-on-T risk.
An R-on-T can still result if an undersensing atrial lead is used in a DDD mode. This occurs when:
- The atrial lead fails to sense a native P wave
- Atrial pacing then occurs during the subsequent (native) QRS
- The (functioning) ventricular lead studiously ignores any sensed electrical activity that occurs during atrial pacing
This is known as a blanking period. - Therefore, the ventricular lead does not register any native ventricular activity that follows from the paced atria
- The (functioning) ventricular lead studiously ignores any sensed electrical activity that occurs during atrial pacing
- After the programmed A-V interval, the ventricular lead paces
- This pacing occurs during repolarisation
Oversensing:
- Inappropriate inhibition of pacing due to misinterpretation of other electrical activity as chamber depolarisation
Sources include:- Diathermy
- Muscle activity
- Shivering
- T-wave
- Other chamber depolarisation
Capture
Capture describes how easily the pacemaker can cause a depolarisation. Capture is:
- Measured in mA
- The capture threshold is the current at which a pacing impulse causes chamber depolarisation
- The capture setting is a current well above the capture threshold
- Good when the capture threshold is low
- Poor when the capture threshold is high
Pacing Codes
A generic 3 or 5 digit code is used to describe pacing modes:
This code is used by the North American Society of Pacing and Electrophysiology (NAPSE), and the British Pacing and Electrophysiology Group (BPEG), and is abbreviated to the NBG (NAPSE/BPEG Generic Code).
- Chambers paced
A, V, D (dual, i.e. both atria and ventricles), O (neither). - Chambers sensed
A, V, D, O. - Response to sensing
- I: Inhibition
Pacemaker is inhibited by a sensed signal. e.g. VVI, ventricular pacing is inhibited by sensed ventricular activity. - T: Trigger
Pacemaker triggered by sensed signal. - D: Inhibition and triggering
Requires a dual-chamber system.
- I: Inhibition
- Rate Response
Only relevant in permanent systems. Describes if the pacemaker changes rate due to sensed change in metabolic demand.- R: Responsive
Responds to either movement or ↑ minute ventilation. - O: No response
- R: Responsive
- Multisite pacing
Are multiple cardiac chambers paced:- O: None
- A: Atria
- V: One or both ventricles
- D°: Any combination of ventricles
Pacing Modes
Mode | Action | Indications | Contraindications |
---|---|---|---|
AAI |
|
|
|
VVI |
|
|
|
DDD |
|
|
|
DDI |
|
|
|
VOO |
|
|
|
DOO |
|
|
|
Temporary Systems
Temporary systems include epicardial and transvenous systems. These systems provide for:
- Setting capture and sensitivity
- Levels should be determined following a pacing check
- Default settings prior to performing a pacing check include:
- Output: 10-15mA
- Sensitivity:
- Rate: ~10 higher than native rate
Particularly prior to pacing check, to ↓ risk of R-on-T.
- Performing rapid atrial pacing
Pacing Check
Full pacing checks cannot be performed if:
- The underlying rate is inadequate:
- Symptomatic hypotension
- MAP <50mmHg
- Temperature <36°C
- Hypokalaemia
K+ <4.5mmol/L.
Temporary epicardial systems should be checked frequently (at least once per nursing shift) to ensure capture and sensitivity thresholds and safe and functional. The full check requires:
- Preparation
- ECG and BP monitoring
- Explain to the awake patient that they may feel pre-syncopal
- Sensitivity testing
Test of the pacemakers detection of native cardiac output.- Set the pacing rate below the native rhythm
- 20bpm below a regular rhythm
- 30bpm below an irregular rhythm
- Set the output of both leads to 0.1mA
This activates both chambers but ensures that R on T will not occur as the output is too low to achieve capture. No pacing spikes will be visible on the ECG. - Check ventricular sensitivity threshold
- Start at the lowest value (0.8mV for the ventricular lead)
- ↑ Sensitivity (mV) until the sensing light is no longer flashing, and the pacing light flashes instead. Then ↓ sensitivity until all native beats are sensed. This is the sensing threshold. The sensitivity setting of the pacemaker is set based upon this threshold.
- Check atrial sensitivity
As ventricular sensitivity. - Check ventricular capture threshold
- Set the pacing rate 10bpm above the native rhythm
- ↑ Output from 0.1mA until 1:1 capture is achieved
This is the capture threshold. The capture setting should be set to the greater of double the threshold or 5mA, whichever is greater.
- Set the pacing rate 10bpm above the native rhythm
- Check atrial capture
As ventricular capture.
- Set the pacing rate below the native rhythm
Lead | Threshold | Setting |
---|---|---|
Atrial | <1mv | Avoid as backup mode. Inappropriate in ward. |
1-4mV | 0.5mV | |
>5mV | 1mV | |
Not checked | 0.5mV. Inappropriate in ward. | |
Ventricular | <2mV | Leave pacing off. |
2-6mV | 1mV | |
7-10mV | 2mV | |
>12mV | 3mV |
Rapid Atrial Pacing
Rapid atrial pacing is a technique for reversion of some atrial tachyarrhythmias by pacing the atria at 20-30bpm faster than the native atrial rate (usually at 260-340bpm).
Rapid atrial pacing is also known as atrial overdrive pacing.
Revertable Rhythms | Non-revertable Rhythms |
---|---|
|
|
Steps:
- Ensure accurate invasive arterial monitoring
- Apply defibrillation pads
In case of precipitation of a malignant arrhythmia. - Stages:
- Set rapid atrial pacing rate ~20bpm below the flutter rate
- Commence rapid atrial pacing, ↑ rate as you do so until at 20-30bpm above the flutter rate
Trial for periods up to 20s.
- If effective, SR should be restored following the rapid pacing period
Troubleshooting
Pacing problems include:
- Failure to pace
- Failure to capture
- Irregular pacing
- Malignant arrhythmias
- Pacemaker-mediated tachycardia
- Crosstalk
Failure to Pace
This can be an emergency in the pacing dependent patient. If unstable:
- Ensure the box is connected
- Press the emergency button
Sets DOO at maximal output (usually 20mA atria, 25mA ventricular). - Connect the pacing box directly to the pacing lead (bypassing the connecting leads)
- Replace the pacemaker
- Institute transcutaneous pacing
- Place a transvenous wire
Pacemaker should deliver a pacing impulse, however:
- No pacing spike is seen
- Chamber capture does not occur
The chamber rate is less than the set rate.
Causes and management include:
- Wire disconnection or migration
- Box is on, and mode is appropriate
- Box is connected to cables
- Wires are connected to the cables
Consider replacing cables or cleaning connections. - Wire location at the skin
- Oversensing
- Sensitivity check
- Control other sources of electrical interference
- Crosstalk
- Discuss with electrophysiology
- Disable less-effective system
Failure to Capture
This can be an emergency in the pacing dependent patient. If unstable, follow the advice for failure to pace.
Pacemaker delivers a pacing impulse but capture does not occur.
Causes and management include:
- Wire disconnection and migration
- Box is on, and mode is appropriate
- Box is connected to cables
- Wires are connected to the cables
Consider replacing cables or cleaning connections. - Wire location at the skin
- ↑ Electrical resistance
- Check capture threshold
- Consider echocardiography
Potential for pericardial collection to ↑ capture thresholds.
- AF
Prevents atrial capture.- Rhythm control
Irregular Pacing
Pacing spikes occurring irregularly, or only irregularly capture the chamber.
Causes and management include:
- Native rate similar to set rate
- Assess underlying rate by ↓ set rate
- Assess for respiratory change in native R-R interval
- Set rate either well above or below the native rate
- Tracking atrial arrhythmia
- Control arrhythmia
- Cardioversion
- Change to non A-V sequential mode
- VVI
- DDI
- Control arrhythmia
- Undersensing
- Check sensitivity
- Disable affected lead if undersensing
- Oversensing
- Sensitivity check
- Control other sources of electrical interference
Malignant Arrhythmia
Pacing system delivers an impulse during a vulnerable part of ventricular repolarisation, resulting in polymorphic VT or VF. Causes and management include:
- Undersensing of atrial or ventricular leads
- Commence CALS
- Pause pacemaker
- Defibrillation
- Following ROSC:
- Repeat sensitivity check
- Review telemetry
Pacemaker-mediated Tachycardia
Broad complex tachycardia occurring just below the upper ventricular rate of the pacing system. Pacemaker-mediated tachycardia:
- Occurs when there is both:
- A-V sequential pacing mode (e.g. DDD)
i.e. A mode with both:- Atrial sensing
- Ventricular pacing
- Slow retrograde AV nodal conduction or accessory pathway
- A-V sequential pacing mode (e.g. DDD)
- Results from:
- Ectopic (or paced) ventricular depolarisation
Ectopy required so that the AV node is not in a refractory state. - Retrograde conduction via the AV node into the atria
- Atrial activity (via retrograde conduction) is sensed, resulting in delivery of a ventricular impulse after the programmed A-V interval
- Ectopic (or paced) ventricular depolarisation
Many pacing systems include a PVARP, which prevents sensing of an atrial impulse shortly after delivery of a ventricular one in order to ↓ risk of pacemaker-mediated tachycardia.
Therefore, for this to occur the retrograde AV nodal conduction needs to be slower than the PVARP so it is registered when the impulse reaches the atria.
Pacemaker-mediated tachycardia may occur with permanent systems but is usually self-terminated, as the system recognises the arrhythmia and defaults to a DDI mode.
Management:
- Pause pacing
This will terminate a pacemaker-mediated rhythm. - ↑ PVARP
Will be limited by the set A-V interval and the set rate. - Change to DDI or VVI
Crosstalk
Crosstalk occurs when there are two pacing systems in use. Crosstalk:
- Usually occurs when there is a temporary epicardial system and a permanent system in place
- May result in failure to pace if:
- The faster system fails to capture the chamber it is pacing
- The slower system senses the attempt to the faster system, and inhibits its own pacing
References
- Bersten, A. D., & Handy, J. M. (2018). Oh’s Intensive Care Manual. Elsevier Gezondheidszorg.