Impella
Catheter-based microaxial flow pump indicated in cardiogenic shock secondary to LV failure, which provides:
- Left ventricular unloading
- ↓ EDV
- ↓ EDP
- ↓ Wall tension
- ↑ Aortic root pressure
- ↑ Coronary perfusion
- ↑ CO
This is in strong contradistinction to VA, which provides systemic perfusion at the expense of ↑ LV afterload and ↑ LV wall tension.
Indications
- Cardiogenic shock secondary to LV failure
Contraindications
- Mechanical aortic valve
- LV thrombus
- Peripheral vascular disease
- Right ventricular failure
Anatomy
Equipment
- Impella catheter
From proximal to distal, consists of:- Blood inlet
- Axial flow pump
Rotational speed is directly proportional to flow. - Radio-opaque strip
Aids placement, as this line should straddle the aortic valve. - Blood outlet
- Optical pressure sensor
- Suture hub
- Automated Impella Controller
Provides:- Interface for monitoring and control
- Display
- Aortic root pressure
- LV pressure
- Flow control
Adjust between ordinal P-levels. - Current calculated flow and P-level
Includes maximal systolic and minimal diastolic flow rate.
- Purge system
Adjusted to maintain a pressure level of 300-1100mmHg; at a rate of 2-30mL/hr. - Battery backup
60 minutes.
- Purge fluid
Purge fluid provides a pressure head inside the device to prevent blood entering the motor. Fluid options include:- Dextrose 5% with heparin (25U/mL)
- Dextrose 5% with bicarbonate
If concern for bleeding or HITT.
With respect to measurements:
- Aortic pressure is monitored by the optical sensor
- Aortic-LV pressure differential is proportional to the motor current
- LV pressure is calculated by the differential between the aortic and LV pressure
P-level | Mean Flow (L/min) | Revolutions per Minute |
---|---|---|
P-0 | 0 | 0 |
P-1 | 0 | 12,000 |
P-2 | 0 - 1.9 | 17,000 |
P-3 | 1.1 - 2.7 | 20,000 |
P-4 | 1.9 - 3.3 | 22,000 |
P-5 | 2.8 - 3.7 | 24,000 |
P-6 | 3.4 - 4.1 | 26,000 |
P-7 | 3.9 - 4.5 | 28,000 |
P-8 | 4.3 - 4.9 | 30,000 |
P-9 | 5.0 - 5.5 | 33,000 |
Notes on Impella flow:
- Pressure gradient through the device is ~30 - 60 mmHg
- Flow varies at each P-level due to ventricular loading
- The Impella is preload dependent and afterload sensitive
Technique
Insertion
Management
Haemodynamics:
- Maintain correct positioning
Indicated by:- A true difference between the aortic and LV pressure waveform
- Inlet 5cm from aortic valve annulus, with the Impella bend at the annulus
- Cardiac arrest
↓ Impella flow rate during arrest.
Weaning:
- ↓ Impella flow rate by 2 P-level increments
Avoid ↓ flow rate <P-2, until immediately prior to removal. - Monitor trend in MAP and LVEDP
- Goal is maintenance of total cardiac output and cardiac power, with stable MAP and LVEDP
Anticoagulation:
- Maintain ACT 160-180
May be achieved with the heparin in the purge fluid.
Removal:
- Continue support at P-2 until immediately prior to removal
- ↓ Flow to P-1 and then withdraw into the aorta
- ↓ Flow to P-0 and remove device
- Surgical closure of graft site
Complications
- Aortic dislodgement
Withdrawal of the device into the aorta.- Indicated by:
- Overlap of the LV and aortic pressure waveform
- Both waveforms consistent with an arterial pressure wave
- Absence of motor current
- Actions
- Reduce P-level to P2
- Reposition catheter under imaging guidance
- Indicated by:
- Ventricular dislodgement
Advancement of the device into the ventricle.- Indicated by:
- Overlap of the LV and aortic pressure waveform
- Both waveforms consistent with an= ventricular pressure wave
- Absence of motor current
- Actions
- Reduce P-level to P2
- Reposition catheter under:
- Without imaging guidance
Gentle withdrawal until pressure and current differential is seen. - With echocardiographic guidance
- Without imaging guidance
- Indicated by:
- Unknown position
May occur when:- The catheter has migrated, but the pressure waveforms are not definitively aortic or arterial
- Native heart function is too impaired to produce a significant pressure differential
- Suction
- Indicated by:
- Suction alarm
- Negative left ventricular diastolic pressure
- Flow lower than expected
- Hypotension
- ↓ Motor current
- May occur when:
- Incorrect ventricular positioning
- RV failure
- Hypovolaemia
- Actions
- Reduce P-level
- Review volume state
- Check RV function
- Check Impella position
- Indicated by:
- Haemolysis
Always pathological, and indicates catheter obstruction:- Inflow obstruction
- Papillary muscles
- Suck-down
- Pump obstruction
- Fibrin
- Clot
- Outflow obstruction
- AV
- Aortic wall
- Inflow obstruction